HR4000CG Spectrometer: UV-NIR
Our HR4000CG Composite-grating (CG) Spectrometer uses our HC-1 Grating -- a revolutionary and proprietary variable-blazed grating specifically designed to provide full spectral output throughout the 200-1100 nm range, with best efficiency at 200-1050 nm.
The revolutionary HR4000CG is a preconfigured HR4000 Spectrometer with all of the optical bench options already selected for you.
This turn-key system is immediately available to you as off-the shelf.
Priced from $5,774.
|
|
Features
-
Wide range (200-1100 nm) broadband spectrometer
-
High resolution (0.75 nm FWHM)
-
Installed HC-1 composite grating and order-sorting filter
-
Works with SpectraSuite Spectrometer Operating Software, available for
200-1100 nm Range, 0.75 to 1.0 nm Resolution
The HR4000CG Spectrometer utilizes a 5 m entrance slit, proprietary HC-1 grating and patented order-sorting filter to provide a 200-1100 nm wavelength range and excellent optical resolution: 0.75 nm (FWHM). The HR4000CG comes with a 3648-element linear-array CCD detector that not only provides better optical resolution throughout the 200-1100 nm range, but also provides improved sensitivity in the 1070-1100 nm region. The HR-series Spectrometers use the HR?Optical Bench and come with UV Detector Upgrades to enhance the performance of the spectrometers in the UV.
Composite-grating & Variable Order-sorting Filter
All CG-series Spectrometers use the HC-1 grating, which is a revolutionary and proprietary variable blazed grating designed to provide a 200-1100 nm wavelength range. (See the grating efficiency curve for the HC-1 at right.) In addition to this composite grating, each CG-series Spectrometer has an OFLV variable longpass order-sorting filter to eliminate second- and third-order effects. Both the grating and OFLV filter are installed at the time of manufacture and are fixed in place.
Programmable Microcontroller
The HR4000CG has an onboard programmable microcontroller that provides flexibility in controlling the spectrometer and accessories. By using a new 30-pin connector, you can implement numerous operating parameters in the software, including controlling external light sources and retrieving data on external objects. You have access to 10 programmable digital I/Os for interfacing to other equipment; one analog input and one analog output; and a pulse generator for triggering devices.
Spectra Obtained with the HR4000CG-UV-NIR
|
CG-series Spectrometer solar irradiance spectrum. |
Physical |
Dimensions: |
148.6 mm x 104.8 mm x 45.1 mm |
Weight: |
570 g |
Detector |
Detector: |
Toshiba TCD1304AP linear CCD array |
Pixels: |
3636 pixels |
Pixel size: |
8 μm x 200 μm |
Pixel well depth: |
~100,000 electrons |
Sensitivity: |
130 photons/count at 400 nm; 60 photons/count at 600 nm |
Optical Bench |
Design: |
f/4, Symmetrical crossed Czerny-Turner |
Focal length: |
101.6 mm input and output |
Entrance aperture: |
5 m wide slit |
Grating: |
HC-1, 300 lines per mm grating |
OFLV filter: |
OFLV-200-1100 |
UV enhanced window: |
Yes, UV4 |
Fiber optic connector: |
SMA 905 to 0.22 numerical aperture single-strand fiber |
Spectroscopic |
Wavelength range: |
200-1050 nm |
Optical resolution: |
0.75 nm FWHM |
Signal-to-noise ratio: |
300:1 (at full signal) |
A/D resolution: |
14 bit |
Dark noise: |
12 RMS counts |
Dynamic range: |
2 x 10^9 (system); 1300:1 for a single acquisition |
Integration time: |
3.8 milliseconds to 10 seconds |
Stray light: |
<0.05% at 600 nm; <0.10% at 435 nm |
Corrected linearity: |
>99.8% |
Electronics |
Power consumption: |
450 mA @ 5 VDC |
Data transfer speed: |
Full scans into memory every 4 milliseconds with USB 2.0; every 18 milliseconds with USB 1.1 |
Inputs/outputs: |
10 onboard digital user-programmable GPIOs |
Analog channels: |
One 13-bit analog input one 9-bit analog output |
Trigger modes: |
4 modes |
Strobe functions: |
Yes |
Connector: |
30-pin connector |
Computer |
Operating systems: |
Windows 98/Me/2000/XP, Mac OS X and Linux operating systems when using the USB port |
Computer interfaces: |
USB 2.0 @ 480 Mbps (USB1.1 compatible); RS-232 (2-wire) @ 115.2 K baud |
Peripheral interfaces: |
SPI (3-wire); I2C inter-integrated circuit |
Pricing: |
|
|