HR2000+ES High-resolution Spectrometer
for Enhanced Sensitivity
The HR2000+ES Spectrometer is a preconfigured system that integrates a high-resolution optical bench, a powerful 2-MHz analog-to-digital (A/D) converter, programmable electronics, a 2048-element CCD-array detector, and a high-speed USB 2.0 port. This innovative combination produces and extremely fast spectrometer and provides resolution to ~1.33 nm (FWHM).
The HR2000+ES has a slightly larger entrance slit (10 microns) and an L2 quartz collection lens to allow for extended measurement applications including:
- Reflection (emissive and reflective color)
- Absorbance
- Irradiance
- Transmission
The HR2000+ES allows users to capture and store a full spectrum into memory every millisecond (that's 1,000 full spectra every second) when the spectrometer is interfaced to a computer via a USB 2.0 port. The HR2000+ES is perfect for chemical, biochemical and other applications where fast reactions need to be monitored and fine spectral features need to be resolved.
- 1,000 Full Spectra/Second
- Programmable Microcontroller
- High-resolution Applications
- HR2000+ES Sample Spectrum
- High-resolution "HR" Optical Bench
- Plug and Play Operation
- Configuration
- Pricing
- Specifications
1,000 Full Spectra/Second
The HR2000+ES utilizes an onboard, 2-MHz A/D converter, which allows you to capture and transfer one full spectrum into memory every millisecond when the spectrometer is interfaced to a PC via the USB port.
Programmable Microcontroller
The HR2000+ES has an onboard programmable microcontroller that provides flexibility in controlling the spectrometer and accessories. Through a new 30-pin connector, you can implement all operating parameters in the software, such as controlling external light sources, creating processes and routines and retrieving data from external devices. The HR2000+ES gives you access to 10 user-programmable digital I/Os for interfacing to other equipment; one analog input and one analog output; and a pulse generator for triggering other devices. (Programming the I/Os requires SpectraSuite Spectroscopy Operating Software.)
High-resolution Applications
The HR2000+ES is ideal for applications where fast reactions need to be monitored and high resolution is necessary, such as protein dynamics.
HR2000+ES Sample Spectrum
|
|
This power-up data for our LS-1 Tungsten Halogen Light Source was taken by an HR2000+ at 2-millisecond intervals. The graph shows the tremendous amount of data generated with the HR2000+s acquisition rate speed of 1000 spectra per second. |
High-resolution "HR" Optical Bench
The exceptional ~1.33 nm optical resolution of the HR2000+ES is achieved by using the Ocean Optics High-resolution Optical Bench ?a 25.4-mm diameter, 101.6-mm focal length, symmetrical crossed Czerny-Turner optical design.
Plug and Play Operation
The HR2000+ES interfaces to a computer via USB 2.0 or RS-232 serial port. When using the serial port, the HR2000+ES requires a single 5-volt power supply (not included). Data unique to each spectrometer are programmed into a memory chip on the HR2000+; SpectraSuite Spectrometer Operating Software reads these values for easy setup and hot swapping among computers.
Configuration
Equipment |
Details |
Spectrometer |
HR2000+ |
Grating |
HC1 |
Detector |
DET2B-200-1100 |
Slit |
SLIT-10 |
Collection Lens |
L2 |
Pricing
Item |
Description |
Price |
HR2000+ES |
HR2000+ES User-configured Spectrometer. system. Includes SpectraSuite and an ASP. |
$6044 |
Specifications
Physical
|
Dimensions: |
148.6 mm x 104.8 mm x 45.1 mm |
Weight: |
570 grams |
Detector |
Detector:
|
Sony ILX511B linear silicon CCD-array
|
Detector range: |
200-1100 nm |
Pixels:
|
2048 pixels
|
Pixel size:
|
14 μm x 200 μm
|
Pixel well depth:
|
~62,500 electrons
|
Sensitivity: |
75 photons/count at 400 nm; 41 photons/count at 600 nm |
Optical Bench |
Design: |
f/4, Symmetrical crossed Czerny-Turner |
Focal length: |
101.6 mm input and output |
Entrance aperture: |
10 m wide slit |
Grating: |
HC-1 provides 200-1100 nm range |
Detector collection lens: |
L2 |
OFLV filter: |
OFLV 200-1100 nm |
UV enhanced window: |
UV2 |
Fiber optic connector: |
SMA 905 to 0.22 numerical aperture single-strand optical fiber |
Spectroscopic |
Wavelength range: |
200-1100 nm (Grating dependent) |
Optical resolution: |
~1.33-6.8 nm FWHM |
Signal-to-noise ratio: |
250:1 (at full signal) |
A/D resolution: |
14 bit |
Dark noise: |
12 RMS counts |
Dynamic range: |
2 x 10^8 (system); 1300:1 for a single acquisition |
Integration time: |
1 ms to 65 s (20 s typical max.) |
Stray light: |
<0.05% at 600 nm; <0.10% at 435 nm |
Corrected linearity: |
>99.8% |
Electronics |
Power consumption: |
220 mA @ 5 VDC |
Data transfer speed: |
Full scans to memory every 1 ms with USB 2.0 port, 15 ms with USB 1.1 port |
Inputs/Outputs: |
Yes, 10 onboard digital user-programmable GPIOs |
Analog channels: |
One 13-bit analog input; one 9-bit analog output |
Auto nulling: |
No |
Breakout box compatibility: |
Yes, HR4-BREAKOUT |
Trigger modes: |
4 modes |
Strobe functions: |
Yes |
Gated delay feature: |
No |
Connector: |
30-pin connector |
Computer |
Operating systems: |
Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
Computer interfaces: |
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud |
Peripheral interfaces: |
SPI (3-wire); I2C inter-integrated circuit |
|