|  At a Glance | 
          
          
            |   | 
            NIRQuest512 | 
            NIRQuest512-2.2 | 
            NIRQuest256-2.1 | 
            NIRQuest256-2.5 | 
          
          
            | Detector: | 
            Hamamatsu G9204-512 InGaAs linear array | 
            Hamamatsu G9204-512W InGaAs linear array | 
            Hamamatsu G9206-256 InGaAs linear array | 
            Hamamatsu G9208-256 InGaAs linear array | 
          
          
            | Detector Range: | 
            850-1700 nm | 
            900-2200 nm | 
            900-2100 nm | 
            900-2550 nm | 
          
          
            | Useable range1: | 
            900-1700 nm | 
            900-2200 nm | 
            900-2050 nm | 
            900-2500 nm | 
          
          
            | Pixels: | 
            512 | 
            512 | 
            256 | 
            256 | 
          
          
            | Pixel size: | 
            25 m x 500 m | 
            25 m x 250 m | 
            50 m x 500 m | 
            50 m x 500 m | 
          
          
            | Entrance aperture: | 
            
             25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit) 
             | 
            
             25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit) 
             | 
            25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit) | 
            25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit) | 
          
          
            | Fiber optic connector: | 
            SMA 905 | 
            SMA 905 | 
            SMA 905 | 
            SMA 905 | 
          
          
            | Grating options (standard): | 
            Grating NIR3, 150 l/mm, 900-1700 nm | 
            Grating NIR2, 100 l/mm, 900-2050 nm | 
            Grating NIR2, 100 l/mm, 900-2050 nm | 
            Grating NIR1, 150 l/mm, 1075-2500 nm | 
          
          
            | Grating options (custom): | 
            Gratings NIR10, NIR11, NIR12, NIR13 and NIR14 | 
            Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 | 
            Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 | 
            Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 | 
          
          
            | Longpass filter2: | 
            OF1-RG830 longpass NIR filter (optional) | 
            OF1-RG830 longpass NIR filter (optional) | 
            OF1-RG830 longpass NIR filter (optional) | 
            OF1-RG830 longpass NIR filter (optional) | 
          
          
            | Wavelength range: | 
            900-1700 nm  
            w/Grating NIR3 | 
            900-2150 nm  
            w/Grating NIR2 | 
            900-2050 nm  
            w/Grating NIR2 | 
            900-2500 nm  
            w/Grating NIR1 | 
          
          
            | Optical resolution (FWHM)3: | 
            ~3.1 nm w/25 m slit | 
            ~4.6 nm w/25 m slit | 
            ~7.6 nm w/25 m slit | 
            ~ 9.5 nm w/25 m slit | 
          
          
            | Signal-to-noise ratio at full signal4: | 
            >15000:1 @ 100 ms integration 
            >13000:1 @ 1000 ms integration | 
            >10000:1 @ 100 ms integration | 
            10000:1 @ 100 ms integration | 
            7500:1 @ 10 ms integration | 
          
          
            | A/D resolution: | 
            16-bit | 
            16-bit | 
            16-bit | 
            16-bit | 
          
          
            | Dark noise: | 
            6 RMS counts @ 100 ms 
            12 RMS counts @  
            1000 ms | 
            6 RMS counts @ 100 ms 
            12 RMS counts @  
            250 ms | 
            6 RMS counts @ 100 ms 
            12 RMS counts @ 250 ms | 
            8 RMS counts @ 10 ms 
            12 counts RMS @ 30 ms | 
          
          
            | Dynamic range: | 
            15 x 10^6 (system); 5000:1 for a single acquisition | 
            15 x 10^6 (system); 1000:1 for a single acquisition | 
            5 x 10^6 (system); 5000:1 for a single acquisition | 
            3  x 10^5  (system); 5000:1 for a single acquisition | 
          
          
            | Integration time5: | 
            1 ms-10 seconds | 
            1 ms-2000 ms | 
            1 ms-2000 ms | 
            1 ms-60 ms | 
          
          
            | Corrected linearity: | 
            >99.8% | 
            >99.8% | 
            >99.8% | 
            >99.8% | 
          
          
            | Computer interfaces: | 
            
             USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud 
             | 
          
          
            | Wireless (Wi-Fi) interface: | 
            
             Yes, with Remora adapter 
             | 
          
          
            | Wi-Fi range: | 
            
             25 meters in free space 
             | 
          
          
            | 
              
            1 Useable range?is defined in the context of the NIRQuest models detector response and its typical grating response. For example, the 512-element detector has response at 850 nm, but grating response begins at 900 nm. The G9206 256-element detector response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 C. The G9208 256-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm. 
            2 Other filter options are available for order-sorting in the NIRQuest256-2.1 and NIRQuest256-2.5. Please contact an Applications Scientist for details. 
            3 Optical resolution (FWHM) depends on grating and slit selection. 
            4 SNR will decrease at longer integration times. 
             
            5 Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level rises to half of full scale. 
             |